- Bring the pouched test device to room temperature(15-30℃) prior to testing. Do not open pouch until ready to perform the assay.
- Remove the test device from the sealed pouch. Lay it on a flat, clean and dry surface.
- Use the pipette to draw and slowly add 1 drop of whole blood/serum/plasma to the sample well.
- Hold the buffer vertically and add 1 drop to the sample well. /If using a pipette, change a new one to avoid cross-contamination. Draw and transfer 2 drops of buffer to the sample well.
- Interpret test results within 10-15 minutes. Do not interpret after 20 minutes.
INTERPRETATION OF RESULTS:
Positive:
Two red lines are visible in the result window. The intensity of the test line may be weaker or darker than that of the control line. This still means a positive result.
Negative:
The control line appears in the result window, but the test line is not visible.
Invalid:
If the control line does not appear in the result window, the test results are INVALID regardless of the presence or absence of the line in the test region.
NOTE: Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for the control line failure. Review the procedure and repeat the test with a new device. If problem persists, please contact your local distributor.
SPECIFICATION:
Description | Specimen | Format | Catalog No. | Sensitivity | Package Size | Strip Width | Certificate |
(TP)Anti-Syphilis Test | Serum/Plasma | Strip | C20-11 | 98.60% | 50 Tests | 4.0 mm | CE |
Cassette | C20-20 | 25 Tests | 4.0 mm | ||||
whole blood | Cassette | C20-21 | 25 Tests | 4.0 mm |